Please use this identifier to cite or link to this item: http://imsear.hellis.org/handle/123456789/72727
Title: Validation of ESR analyzer using Westergren ESR method.
Authors: Sikka, Meera
Tandon, Rajesh
Rusia, Usha
Madan, Nishi
Issue Date: 22-Jul-2007
Citation: Sikka M, Tandon R, Rusia U, Madan N. Validation of ESR analyzer using Westergren ESR method. Indian Journal of Pathology & Microbiology. 2007 Jul; 50(3): 634-5
Abstract: Erythrocyte sedimentation rate (ESR) is one of the most frequently ordered laboratory test. ESR analyzers were developed to provide a quick and efficient measure of ESR. We compared the results of ESR obtained by an ESR analyzer with those by the Westergren method in a group of 75 patients Linear regression analysis showed a good correlation between the two results (r = 0.818, p < 0.01). The intra class correlation was 0.82. The analyzer method had the advantages of safety, decreased technician time and improved patient care by providing quick results.