Title: Can ocular torsion be measured using the slitlamp biomicroscope?
Authors: Kothari, Mihir T
Venkatesan, G
Shah, Jugal P
Kothari, Kulin
Nirmalan, Praveen K
Issue Date: 15-Mar-2005
Citation: Kothari MT, Venkatesan G, Shah JP, Kothari K, Nirmalan PK. Can ocular torsion be measured using the slitlamp biomicroscope? Indian Journal of Ophthalmology. 2005 Mar; 53(1): 43-7
Abstract: PURPOSE: To evaluate and compare a stereo-biomicroscopic technique of ocular torsion measurement with fundus photographic technique. METHODS: This prospective masked observational study included 72 eyes of 36 consecutive subjects with normal ocular motility. The disc-foveal angle was measured by two masked investigators using a slitlamp stereo-biomicroscopic and fundus photographic technique. RESULTS: The mean age was 13.7 years (6 - 44 years), and 15 (41.6%) were males. The average torsion using the slitlamp technique was 5.5 +/- 3.3 degrees and 6.1 +/- 4.3 degrees with fundus photographic technique. Clinical agreement between the two techniques for 5 degree and 6 degree torsion was 81% and 90% respectively. The average difference between the two techniques was 3.0 +/- 2.6 degrees . The mean inter-ocular difference was 3.3 +/- 2.7 degrees (SD, 2.77) and 5.5 +/- 4.6 degrees using the slitlamp and fundus photographic technique respectively. The Pearson correlation coefficient (r2) was 0.5. CONCLUSION: The fundus photographic technique remains the gold standard test for the measurement of ocular torsion. The slitlamp biomicroscopic method can be useful where a fundus camera is unavailable. Further studies are required to test the repeatability of these tests.
URI: http://imsear.hellis.org/handle/123456789/71530
Appears in Collections:Indian Journal of Ophthalmology

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.